Question | Answer |
How many functions does the WAM perform? | 7 fully automatic functions including Ref/Ker/Topography/Pupilometry/Tonometry/Pachymetry/Wavefront HOA & LOA Analysis + retro-illumination |
how long does it take to run all 7 functions? | 90 seconds to automatically complete all 7 functions for both eyes |
Does it need a PC or software? | No PC needed, it is built into the machine. Viewing can be done through the network with remote access. |
Can we extract reports for the EHR? | Reports can be exported as a PDF for import into an EHR |
How does it measure Topography | 24 Placido Rings measuring 6144 points |
How does it measure Wavefront | Shack-Hartmann Wavefront Sensor with 100,000 points analyzed. |
how many Viewer licenses delivered with WAM700? | Up to 25 viewers |
Is a PC required? | A PC is not required unless you want to view the patient information remotely. |
Do I need the Internet for the WAM700? | Only if you will export XML, PDF or access remotely. |
Can I connect a TV or PC monitor to WAM700? | Yes, via VGA cable or VGA to HDMI converter |
Is WAM700 DICOM compatible? | No, there it not DICOM possibilities. |
Do we need USB3 port in the PC? | No, USB 3.0 is not a requirement |
Can we do MEIBO analysis with WAM700 | this is not possible. |
What is the resolution of the images? | 1080 x 1920 Pixels |
Can I do a video? | No, only pictures |
Can I change the contrast/gamma/brightness of the image? | Only can be done on the pachy images |
What is the minimum pupil size? | 2.0 mm |
Do I need to drill my table to mount the WAM700? | No, drilling is not needed. |
Do we have Soft lenses library into WAM700? | No, only RGPs |
Do we have Scleral lenses library into WAM700? | No, only RGPs |
Can we do TEAR FILM or BUT with WAM700? | Is not possible. |
How many Placido rings it has? | 24 |
How many points are analyzed? | 6144 |
How is this compared with Pentacam form Oculus? | TBC |
Can I measure the elevation of both the anterior and posterior cornea? | No on both, only on anterior. |
Does the software has cases of corneal pathologies such as keratoconus, the corneal apex is decentered off the visual axis? | Some cataract and HOA/LOA cases |
Can I do multiple maps display? | No |
Can I do OD/OS compare display? | Yes |
Can I do an overlay of the image for OD/OS ? | No |
Can I do full study of the pupil diameters? | Yes |
Can I simulate NaFl (fluorescein)? | Yes |
Can I do picture and videos of the Anterior Segment? | No |
Does it include features for Keratoconus assessments? | No |
Can I do printouts of the display? | Yes |
Do I have Scheimpflug technology? | Yes, but static |
Anaeyes has corneal ectasia monitoring and treatment? | No, it does not have any monitoring module. |
Can I do keratoconus progression on patients? | Yes is possible. |
Is WAM700 an Anterior Segment Optical Coherence Tomography (AS-OCT) | No, is not an AS-OCT |