Procedure to activate Randomizing - APH550 and CS550

Procedure to activate Randomizing - APH550 and CS550


Do not attend to update your units without phone support.
Ignoring this recommendation could lead you to brick your units and void any warranties.
The procedure has been designed as a guideline tool.

Step 1
Mention that you desire to activate the Randomizing Function between APH550 and CS550

Step 2 
Download the files attached below.

Step 3 


Connect the cable provided into the this port labeled PC.
The other end connect it into any USB port on your PC or Laptop. 

Step 4
When is time the customer support representative will instruct you review this graphics.


Displays for reference.




Communication speed
It is set as 115200bps by pressing the top button and set as 9600bps by pressing the
bottom button. (Set the communication speed same with PC.)

Hyper mode
Enter the Hyper mode by pressing the right button. It is cancelled by pressing the left  button.

Press the XC1 button after setting the communication.
“Please Send MOT File.” is displayed by pressing the XC1 button.

Communication execution
Start the communication by pressing the “Send” button of the application. The
progressing bars of both the application and APH550 are displayed. Turn the power off
after 3 bars of APH550 become 100% and the “Checksum” is displayed.


Step 5


Make sure you have a USB 3.0  THUMB DRIVE.

5.1.  Use the     power switch to turn off the unit.


5.2   When is time the customer support representative will instruct you review this steps and graphics.
Plug the USB THUMB DRIVE in the indicated port.


5.3  Use the     power switch to turn on the unit.
Below is the booting and updating screen sequence.
Splash Logo
Loading system
Update in process
Update completed
Unit is Off
5.4 Let the unit turn on.
5.5 Unit is reading files
5.6 Unit is updating the software
5.7 Unit completed the update. You can turn off the unit. 
Use the power switch.
                   
5.8  Remove the USB THUMB DRIVE







5.9  Use the     power switch to turn on the unit.

5.10  Connect the remote control into the screen unit CS550
Use the USB cable to recharge the remote control
           

5.11  Align the remote control with the CS550 screen.

5.12  On the remote control press and hold  
ENG MODE screen if followed on the remote control and CONFIGURATION on the CS550
   

15.13    Verify the software version is KB11-1





    • Related Articles

    • APH 550 Automatic Phoropter

      Problem Possible Causes The power indicator of the relay box is not turned on. The power cord may not be properly connected. Make sure to connect it securely. The fuse may be blown. If so, replace it to the new one after turning the power off. The ...
    • CS 550 & CSPOLA 600 Acuity System - FAQ

      Question Answer What is the size of screen CS550? 19" What is the size of screen CSPOLA600? 24" POLARIZATION function Can I do glare test? No, glare test is not possible. What is the min-max length for my exam room? recommend min 10ft - max 18 (21' ...
    • CS 550 & CSPOLA 600 Acuity System

      Problem Possible Causes Suggested Solutions Product does not turn on. Power cable not plugged in. Unplug and replug the power cable to ensure proper fit. No light on power supply Change power supply Product will not respond to remote control. IR ...
    • SL500 & SL550 - Installation Procedure with DS550

      Step 1 Locate the part  ARM005 or 1040X the Installation_Pattern  and TABTOP03 Step 2 Place the Installation_Pattern on the surface of the table and align it with the center. Make sure to flush the line with the edge of the table. Step 3 Use scotch ...
    • Data is not transferring - CBOX

      Issues with CBOX - Data is not transferring Reboot the CBOX and APH.